Parr-instrument Series 4570 Uživatelský manuál Strana 1

Procházejte online nebo si stáhněte Uživatelský manuál pro Měřící nástroje Parr-instrument Series 4570. Parr Instrument Series 4570 User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk

Shrnutí obsahu

Strany 1 - Operating Instruction Manual

397M4570HP/HT Pressure ReactorsOperating Instruction Manual

Strany 2 - TABLE OF CONTENTS

- 10 - 4570 HP/HT Pressure Reactors Parr Instrument Company IDENTIFY THE VALVES Gas Inlet Valve The gas inlet valve is easily identified when the

Strany 3 - PREFACE

- 11 - 4570 HP/HT Pressure Reactors Parr Instrument Company OTHER VESSEL HEAD FITTINGS Safety Rupture Disc There is a safety rupture disc attache

Strany 4 - PREFACE (Continued)

- 12 - 4570 HP/HT Pressure Reactors Parr Instrument Company HOW TO USE THE VESSEL Fixed Head Vessels: First, lower the heater and push it to

Strany 5

- 13 - 4570 HP/HT Pressure Reactors Parr Instrument Company HOW TO USE THE VESSEL (Continued) Pressurizing the Vessel Check all valves carefully b

Strany 6 - INSTALLATION

- 14 - 4570 HP/HT Pressure Reactors Parr Instrument Company ACCESSORIES Internal Cooling Coil A serpentine cooling coil is mounted to the under

Strany 7 - ASSEMBLE THE REACTOR

- 15 - 4570 HP/HT Pressure Reactors Parr Instrument Company ACCESSORIES (continued) Air Motor Variable stirring speeds from 300 to 2000 rpm with

Strany 8

- 16 - 4570 HP/HT Pressure Reactors Parr Instrument Company The flow control valves have been preset for optimum performance. Both the speed and lif

Strany 9

- 17 - 4570 HP/HT Pressure Reactors Parr Instrument Company EXPLOSION PROOF OPERATION If the local safety code requires that equipment installed

Strany 10 - IDENTIFY THE VALVES

- 18 - 4570 HP/HT Pressure Reactors Parr Instrument Company PERIODIC PRESSURE TESTS Each cylinder used in a Parr stirred reactor is tested under hy

Strany 11 - OTHER VESSEL HEAD FITTINGS

- 19 - 4570 HP/HT Pressure Reactors Parr Instrument Company GENERAL MAINTENANCE NOTES 1. Periodically inspect all electrical wiring and pressure co

Strany 12 - HOW TO USE THE VESSEL

- 2 - 4570 HP/HT Pressure Reactors Parr Instrument Company TABLE OF CONTENTS Related Instructions...2 Customer Service...

Strany 13

- 20 - 4570 HP/HT Pressure Reactors Parr Instrument Company REACTION VESSEL PARTS LIST Consult the itemized list for your reactor, provided along

Strany 14 - ACCESSORIES

- 21 - 4570 HP/HT Pressure Reactors Parr Instrument Company Part No. Description Code Stirrer Shafts* - Standard one piece A1120HC (LD) / A1180HC

Strany 15 - ACCESSORIES (continued)

- 22 - 4570 HP/HT Pressure Reactors Parr Instrument Company Part No. Description Code External Fittings Gage Port 208HC15 Gage adapter

Strany 16

- 23 - 4570 HP/HT Pressure Reactors Parr Instrument Company Model 4571 1000 mL Removable Reactor

Strany 17 - OPERATION

- 24 - 4570 HP/HT Pressure Reactors Parr Instrument Company Overarm Parts List Part No. Description Motor A388E6 Air Motor A388EES Motor

Strany 18 - PERIODIC PRESSURE TESTS

- 25 - 4570 HP/HT Pressure Reactors Parr Instrument Company Overarm Assembly

Strany 19 - GENERAL MAINTENANCE NOTES

- 26 - 4570 HP/HT Pressure Reactors Parr Instrument Company Vessel Heater Stand & Parts List For Removable Vessels Part No. Description

Strany 20 - REACTION VESSEL PARTS LIST

- 27 - 4570 HP/HT Pressure Reactors Parr Instrument Company 4570 Vessel Stand2350HC9 Motor encasement 2565HC Lower back plate 2566HC Upper back plat

Strany 22

- 3 - 4570 HP/HT Pressure Reactors Parr Instrument Company PREFACE Scope These instructions describe the installation, operation and maintenance

Strany 23

- 4 - 4570 HP/HT Pressure Reactors Parr Instrument Company PREFACE (Continued) Explanation of Symbols II On position, full power heater switch I

Strany 24 - Overarm Parts List

- 5 - 4570 HP/HT Pressure Reactors Parr Instrument Company PREFACE (Continued) Intended Usage This system has been designed for use as a high pre

Strany 25 - Overarm Assembly

- 6 - 4570 HP/HT Pressure Reactors Parr Instrument Company INSTALLATION Pressure and Temperature Limits The working pressure and temperature at w

Strany 26 - Parr Instrument Company

- 7 - 4570 HP/HT Pressure Reactors Parr Instrument Company ASSEMBLE THE REACTOR These reactors require at least 10 sq. ft. of workspace in a well-ve

Strany 27 - 4570 Vessel Stand

- 8 - 4570 HP/HT Pressure Reactors Parr Instrument Company ASSEMBLE THE REACTOR (Continued) The pneumatic package includes a pressure regulator wit

Strany 28 - Revision 03/14/13

- 9 - 4570 HP/HT Pressure Reactors Parr Instrument Company ASSEMBLE THE REACTOR (Continued) 6. Connect the thermocouple and extension wire to both

Komentáře k této Příručce

Žádné komentáře