Micro Bench Top Reactors
Operating Instruction Manual
Zobrazit stránku 0
1 2 3 4 5 6 ... 27 28

Shrnutí obsahu

Strany 1 - Operating Instruction Manual

399M4590 Micro Bench Top ReactorsOperating Instruction Manual

Strany 2

4590 Micro ReactorsParr Instrument Company10IDENTIFY THE VALVESGas Inlet ValveThe gas inlet valve is easily identified when the vessel is open by notin

Strany 3

4590 Micro Reactorswww.parrinst.com11OTHER VESSEL HEAD FITTINGSSafety Rupture DiscThere is a safety rupture disc attached to the head that is intended

Strany 4

4590 Micro ReactorsParr Instrument Company12Air MotorVariable stirring speeds from 100 to 2000 rpm with no spark hazard can be obtained by replacing t

Strany 5

Noteswww.parrinst.com13

Strany 6

4590 Micro ReactorsParr Instrument Company14HOW TO USE THE VESSELRemovable VesselsAlways remove removable vessels from the support stand before attem

Strany 7

4590 Micro Reactorswww.parrinst.com15Gas ConnectionsFor a gas connection to the vessel, use the A495HC pressure hose (A506HC in high pressure reactors

Strany 8

4590 Micro ReactorsParr Instrument Company16MAINTENANCEGeneral Maintenance Notes1. Periodically inspect all electrical wiring and pres-sure connectio

Strany 9

4590 Micro Reactorswww.parrinst.com17Periodic Pressure TestsEach cylinder used in a Parr stirred reactor is tested under hydrostatic pressure to the

Strany 10 - 4590 Micro Reactors

4590 Micro ReactorsParr Instrument Company18SERIES 4590 VESSEL PARTS LISTSStandard Vessel Fittings (3000 psig / 350 °C)Standard Reaction Vessel Parts

Strany 11

4590 Micro Reactorswww.parrinst.com19Heads*Part No. Description Code2432HC Head RV FG2902HC Head RV OR2772HC Head FH FG2774HC Head FH OR1432HC Head SC

Strany 13

4590 Micro ReactorsParr Instrument Company20R/D ASSEMBLYW/526HC_STYLE RD302VBADFEMALECONNA138CAMALE CONNA2067HC_STYLETHERMOWELLA1357E_STYLET/CA451VBVA

Strany 14

4590 Micro Reactorswww.parrinst.com21High Pressure Reaction Vessel Parts ListConsult the itemized list for your reactor provided along with this manua

Strany 15

4590 Micro ReactorsParr Instrument Company22THERM429HC2KLGASKET, GRAFOIL40HCADAPTER, A SKT X1/4 NPTFVALVE, STRAIGHTPLUG, HEAD 1/4 NPTM1443HC2NIPPLE, H

Strany 16

4590 Micro Reactorswww.parrinst.com233340HC2HEADHigh Pressure / High Temperature Reaction Vessel Parts ListConsult the itemized list for your reactor

Strany 17

4590 Micro ReactorsParr Instrument Company24Vessel Heaters (Standard Stand Configuration)Part No. DescriptionA3240HC9EB Heater Assembly, TW, 700W, 115V

Strany 18

4590 Micro Reactorswww.parrinst.com25Support Stands4591/4592 Micro Reactor with 1/8 hp Variable Speed Motor1.49 (DOWN POSITION)2.05 (UP POSITION)SW51F

Strany 19

4590 Micro ReactorsParr Instrument Company264591/4592 Micro Reactor with Variable Speed 1/4 HP and Explosion Proof Motors(SEE MOTOR TABLE)56C MOUNT MO

Strany 20

4590 Micro Reactorswww.parrinst.com274593 Micro Reactor with Fixed Head

Strany 21

399M R06 01/22/15

Strany 22

4590 Micro Reactorswww.parrinst.com3PREFACE 4Scope 4Related Instructions 4Intended Usage 4Safety Information 4General Specifications 4Elect

Strany 23

4590 Micro ReactorsParr Instrument Company4PREFACEScopeThese instructions describe the installation, opera-tion and maintenance of Parr Series 4590 Be

Strany 24

4590 Micro Reactorswww.parrinst.com5Explanation of SymbolsII On position, full power heater switchI On position, half power heater switchO Off Positio

Strany 25

Parr Instrument Company64590 Micro ReactorsUser’s ResponsibilityAll Parr Reactors and pressure vessels are designed and manufactured with great care t

Strany 26

4590 Micro Reactorswww.parrinst.com7Parr Series 4590 Micro Reactors are furnished with two structural options in addition to the size, pressure range,

Strany 27

Parr Instrument Company84590 Micro ReactorsINSTALLATIONPressure and Temperature LimitsThe working pressure and temperature at which any reactor or pre

Strany 28 - 399M R06 01/22/15

4590 Micro Reactorswww.parrinst.com9Assemble The ReactorThese reactors require at least 10 sq. ft. of workspace in a well-ventilated area with conven