74
6200Calorimeter O
eratin
Instruction Manual
Table D-2
6200 Calorimeter Run Data Template
Field Description
SampleID Field will hold ID up to sixteen characters in length
Timestamp MM/DD/YY HH:mm:ss
Mode 0 = determination, 1 = standardization
Method 0 = equilibrium, 1 = dynamic
State 0 = preweigh, 1 = preliminary, 2 = final
Units 0 = MJ/kg, 1 = Btu/lb, 2 = cal/g, 3 = J/kg, 4 = other
UnitMultIfOther unit multiplier in effect at time of report
BombID [1-4]
BombEE bomb energy equivalent
SampleWt sample weight
SpikeWt spike weight
Fuse fuse value
FuseFinal fuse value is final
Acid acid value
AcidFinal acid value is final
Sulfur sulfur value
SulfurFinal sulfur value is final
Hydrogen hydrogen value (net calc option)
HydrogenFinal hydrogen value is final (net calc option)
Moisture moisture value (dry calc option)
MoistureFinal moisture value is final (dry calc option)
JacketTemp jacket temperature
InitTemp initial temperature
DeltaT temperature rise
HOC gross heat of combustion
NetHOC dry net HOC (if both dry and net calc options enabled)
DryHOC dry gross HOC (if dry calc option enabled)
DryNetHOC dry net HOC (if both dry and net calc options enabled)
D-4
Test data files are named with the following convention.
Test type Filename
Preliminary Standardization <ID>.std.plim.csv
Final Standardization <ID>.std.finl.csv
Preliminary Determination <ID>.det.plim.csv
Final Determination <ID>.det.finl.csv
Pre-weigh <ID>.---.pwgh.csv
APPENDIX D